IR-FA is monochromatic fibre pyrometers which allow to measure temperature in range 200 - 3000°C.
| Type | IR-FAI | IR-FAS | |
|---|---|---|---|
| Measuring system | Single color type | ||
| Element | InGaAs | Si | |
| Wavelength | monochromatic | ||
| Accuracy ratings | Lower than 1000°C: ±5°C 1000°C to 1500°C: ±0.5% of measured value 1500°C to 2000°C: ±1.0% of measured value Higher than 2000°C: ±2% of measured value |
||
| Repeatability | 0,2°C | ||
| Temperature drift | 0.1°C/°C or 0.015% of measured value, whichever is larger |
||
| Resolution | 0,5°C | ||
| Response time | 0,01 s | ||
| Emissivity | (ratio) compensation: 1,999 ... 0,500 | ||
| Signal modulation | DELAY: Tracing of average value (smoothing) (Modulation ratio: 0.0 to 99.9 seconds, 0.1 second increment) Modulation ratio 0 = REAL PEAK: Tracing of maximum value ( Modulation ratio: 0, 2, 5, 10°C/second ), Modulation ratio 0 = PEAK HOLDi |
||
| Display | LCD 4-digit (Temperature display, parameter display), °C/°F (key selection) | ||
| Analog output | 4 to 20mADC isolated output (load resistance: 500Ω or lower) Accuracy rating: ±0.2% of output range Output resolution: 0.01% of output range (IR─FAC); 0.01% of output range (except IR─FAC) Output scaling: Setting within measuring temperature range Dummy output: Setting within 0 to 100% of analog output |
||
| Contact output | 1-point, High (low) alarm or error signal, Photo-coupler 30VDC Maximum 50mA |
||
| Contact input | 1-point, peak hold reset or sample hold, Dry contact or open collector | ||
| Parameter setting by keys | Operator mode: Settings of emissivity, signal modulation, alarms, etc. Engineering mode: Settings of engineering unit (°C/°F), output scaling, zero, span, reference temperature input for automatic emissivity calculation, output correction, etc., Settings of optional functions |
||
| Computing function | Zero/span adjustment, automatic emissivity (ratio) computation (*2), Output correction | ||
| Self-diagnosis | Thermometer temperature abnormal, Parameter error | ||
| Options | Laser function | Solid state built-in laser unit, 1mW (645nm) or less, Class 2 (not available in high sensitivity type) | |
| Analog input | Input signal: 4 to 20mADC Remote setting of emissivity (ratio) or reference temperature input setting for automatic emissivity (ratio) computation |
||
| Communications IF | RS─485, Transmitting of measured data (down to 1 decimal point), Transmitting/receiving of parameters | ||
| Working temperature | 0...50oC | ||
| Rated power supply | 24VDC (Allowable voltage fluctuation range: 22 to 28VDC), Recommended power unit: IR─ZFEP | ||
| Power consumption | Maximum 3VA | ||
| Connection | Cramp type no screw terminals | ||
| Mounting | DIN rail mounting or wall mounting | ||
| Case material | Resin | ||
| Dimensions | 90x90x60 mm | ||
| Weight | About 0,25 kg | ||
| CE CE | EN55011 Grupa 1 Klasa A, EN50082-2 | ||
|
range Lens assembly
|
||||||||||||||||||||||||||||||||
Dimensions [mm] |
Pyrometer |
![]() |
Fiber without metallic protection tube |
Fiber with metallic protection tube |
|||||||||||||||||||||||
![]() |
![]() |
|||||||||||||||||||||||
| Lens assembly |
Air purge case |
|||||||||||||||||||||||
![]() |
![]() |
|||||||||||||||||||||||
Soczewki0,1,2,345678
|
Soczewki0,1,2,345678
|
|||||||||||||||||||||||
Lens assembly |
|
|||||||||||||||||||||||
![]() |
![]() |
|||||||||||||||||||||||
|
|
|||||||||||||||||||||||
Ar Jūs domina šis produktas? Ar Jums reikia papildomos informacijos ar individualaus pasiūlymo?
tu turi būti prisijungęs
IR-FA is monochromatic fibre pyrometers which allow to measure temperature in range 200 - 3000°C.
| Type | IR-FAI | IR-FAS | |
|---|---|---|---|
| Measuring system | Single color type | ||
| Element | InGaAs | Si | |
| Wavelength | monochromatic | ||
| Accuracy ratings | Lower than 1000°C: ±5°C 1000°C to 1500°C: ±0.5% of measured value 1500°C to 2000°C: ±1.0% of measured value Higher than 2000°C: ±2% of measured value |
||
| Repeatability | 0,2°C | ||
| Temperature drift | 0.1°C/°C or 0.015% of measured value, whichever is larger |
||
| Resolution | 0,5°C | ||
| Response time | 0,01 s | ||
| Emissivity | (ratio) compensation: 1,999 ... 0,500 | ||
| Signal modulation | DELAY: Tracing of average value (smoothing) (Modulation ratio: 0.0 to 99.9 seconds, 0.1 second increment) Modulation ratio 0 = REAL PEAK: Tracing of maximum value ( Modulation ratio: 0, 2, 5, 10°C/second ), Modulation ratio 0 = PEAK HOLDi |
||
| Display | LCD 4-digit (Temperature display, parameter display), °C/°F (key selection) | ||
| Analog output | 4 to 20mADC isolated output (load resistance: 500Ω or lower) Accuracy rating: ±0.2% of output range Output resolution: 0.01% of output range (IR─FAC); 0.01% of output range (except IR─FAC) Output scaling: Setting within measuring temperature range Dummy output: Setting within 0 to 100% of analog output |
||
| Contact output | 1-point, High (low) alarm or error signal, Photo-coupler 30VDC Maximum 50mA |
||
| Contact input | 1-point, peak hold reset or sample hold, Dry contact or open collector | ||
| Parameter setting by keys | Operator mode: Settings of emissivity, signal modulation, alarms, etc. Engineering mode: Settings of engineering unit (°C/°F), output scaling, zero, span, reference temperature input for automatic emissivity calculation, output correction, etc., Settings of optional functions |
||
| Computing function | Zero/span adjustment, automatic emissivity (ratio) computation (*2), Output correction | ||
| Self-diagnosis | Thermometer temperature abnormal, Parameter error | ||
| Options | Laser function | Solid state built-in laser unit, 1mW (645nm) or less, Class 2 (not available in high sensitivity type) | |
| Analog input | Input signal: 4 to 20mADC Remote setting of emissivity (ratio) or reference temperature input setting for automatic emissivity (ratio) computation |
||
| Communications IF | RS─485, Transmitting of measured data (down to 1 decimal point), Transmitting/receiving of parameters | ||
| Working temperature | 0...50oC | ||
| Rated power supply | 24VDC (Allowable voltage fluctuation range: 22 to 28VDC), Recommended power unit: IR─ZFEP | ||
| Power consumption | Maximum 3VA | ||
| Connection | Cramp type no screw terminals | ||
| Mounting | DIN rail mounting or wall mounting | ||
| Case material | Resin | ||
| Dimensions | 90x90x60 mm | ||
| Weight | About 0,25 kg | ||
| CE CE | EN55011 Grupa 1 Klasa A, EN50082-2 | ||
|
range Lens assembly
|
||||||||||||||||||||||||||||||||
Dimensions [mm] |
Pyrometer |
![]() |
Fiber without metallic protection tube |
Fiber with metallic protection tube |
|||||||||||||||||||||||
![]() |
![]() |
|||||||||||||||||||||||
| Lens assembly |
Air purge case |
|||||||||||||||||||||||
![]() |
![]() |
|||||||||||||||||||||||
Soczewki0,1,2,345678
|
Soczewki0,1,2,345678
|
|||||||||||||||||||||||
Lens assembly |
|
|||||||||||||||||||||||
![]() |
![]() |
|||||||||||||||||||||||
|
|
|||||||||||||||||||||||