|  Technical Data: | |
| Relative Humidity | |
| - measuring range | 0 ... 100 % rh | 
| - resolution | 0,5 %rh | 
| - accuracy | ±2 %rh 05„.35°C 25...85 %rh) | 
| ±3 %rh (15...35°C, 5...25 %rh and 85...95 %rh) | |
| - Additional error | ±0,1 % rh/K @t35°C | 
| Temperature | |
| - measuring ranges | see table | 
| - resolution | 0.1°C | 
| -accuracy | 0.5 K @ 25°C ±1 digit | 
| digit(see diagram) | |
| General: | |
| - operating voltage | 5...30 V DC | 
| - output signal: | |
| calHT01/1, calHT01/2 | 2 x 0...1 V (przy RL 3 1 MΩ) | 
| calHT 06/1, calHT 06/2 | digit (compatible with I2C) | 
| - storage temperature | - 65 ... 150°C | 
| - working temperature | - 40 ... 85°C | 
| - power consumption | < 1 mA | 
| Type | Measuring range [°C] | Labelling | 
| calHT01-0100(.R) | 0...100 | Green | 
| calHT01-3070(.R) | -30...70 | Black | 
| calHT01-2080(.R) | -20..80 | Red | 
| calHT01-4060(.R) | -40...60 | Yellow | 
| calHT06-4085(.R) | -40...85 | White | 


|  |  | 
| Temperature accuracy of calHT 01 and calHT 06 series | Temperature accuracy of calHT 01 and calHT 06 series | 

| S | 1 | 0 | 0 | 1 | 0 | 0 | 0 | 0 | A | 0 | 0 | 0 | 0 | C 4 | C 3 | C 2 | C 1 | A | P | 
| R/W | |||||||||||||||||||
 Microcontroller requests data byte according control command and reads out
| S | 1 | 0 | 0 | 1 | 0 | 0 | 0 | 1 | A | D 7 | D 6 | D 5 | D 4 | D 3 | D 2 | D 1 | D 0 | N A | P | 
| S Start | R/W | ||||||||
| P Stop | |||||||||
| A Acknowledge | from sensor | ||||||||
| N/A No Acknowledge | |||||||||
| C 4 | C 3 | C 2 | C 1 | |
|---|---|---|---|---|
| 0 | 0 | 0 | 0 | Read-Out Humidity | 
| 0 | 0 | 1 | 0 | Read-Out Temperature Byte 1 | 
| 0 | 0 | 1 | 1 | Read-Out Temperature Byte 2 | 
| D 7 | D 6 | D | D 4 | D 3 | D 2 | D 1 | D 0 | |
|---|---|---|---|---|---|---|---|---|
| 0/1 | 0/1 | 0/1 | 0/1 | 0/1 | 0/1 | 0/1 | 0/1 | Humidity | 
| 0/1 | 0/1 | 0/1 | 0/1 | 0/1 | 0/1 | 0/1 | 0/1 | Temperature Byte 1 | 
| X | X | X | X | 0/1 | 0/1 | 0/1 | 0/1 | Temperature Byte 2 | 
| Symbol | Parameter | min | max | jednostka | 
|---|---|---|---|---|
| tBUF | idle period between BUS actions | 4,7 | μs | |
| tD:SU | data set-up time | 250 | ns | |
| tD:HD | data hold time | 50 | ns | |
| tST:HD | start hold time | 4 | μs | |
| tST:SU | start set-up time | 4,7 | μs | |
| tSCL:L | SCL„low" time | 4,7 | μs | |
| tSCL:H | SCL „high" time | 4 | μs | |
| fSCL | SCL frequency | 100 | kHz | |
| t r | SDA, SCL LOW/HIGH time | 1 | μs | |
| tf | SDA, SCL HIGH/LOW time | 0,3 | μs | |
| tSTO:SU | stop set-up time | 4 | μs | |
| tSP | pulse width of spikes to be suppressed by input filter | 100 | ns | |
| CL | capacitive load SDA, SCL BUS (internal pull-up 120 kW) | 10 | pF | |
| tMUPD | update for measurement values | 40 | ms | |
| tSMPL | sampling period DA converter | 2 | 6 | ms | 

Instrukcja użytkowania
 Nigdy nie wolno dotykać aktywnej części bardzo czułego czujnika. Należy używać bardzo niskiego napięcia podczas lutowania elementu (temperatura lutowania 240°C, czas lutowania 2 sekundy). Należy oczyścić element pomiarowy delikatnie zdmuchując wszelkie pozostałości z montażu, można również przepłukać element w wodzie destylowanej. Nie wpływa to na charakterystykę czujnika, jednak dokładne pomiary mogą być dokonywane dopiero po całkowitym wysuszeniu elementu. Dotyczy to również sytuacji w której powstanie na elemencie rosa.
| Typ | FE 09/1, FE 09.R/1 FE 09/2, FE09.R/2 | FE 09/4 FE 09.R/4 | 
|---|---|---|
| Zakres działania | 0...100%rh | 0...100%rh | 
| Temperatura działania FE09/X | -60...200°C | -60...200°C | 
| Temperatura działania FE09.R/X | -40...110°C | -40...110°C | 
| Pojemność podstawowa | 135±10pF | 115±15pF | 
| Zmiana pojemności | 0,3±0,05pF/%rh | 0,27±0,08pF/%rh | 
| Dopuszczalne napięcie | maks. 3V bez DC | maks. 3V bez DC | 
| Częstotliwość pomiarowa | 5...200kHz | 5...200kHz | 
| Współczynnik strat | <0,03 (przy 10kHz) | <0,03 (przy 50kHz) | 
| Czas reakcji (T90) | 10s | 10s | 
| Histereza (MR 5...95%rh) | <1,5%rh | <1,5%rh | 
| Liniowość (MR 5...95%rh) | <1,5%rh | <1,5%rh | 
| Stabilność długoterminowa | <1%rh/a | <1%rh/a | 



| Substancja robocza | stężenie | 
|---|---|
| NH3 (tylko na krótki czas) | 100 ppm (2 x MAK) | 
| N2S | 20 ppm (2 x MAK) | 
| SO2 | 3,7 ppm (10 x MAK) | 
| No2 | 1,0 ppm (10 x MAK) | 
| O2 | 0,6 ppm (10 x MAK) | 
| Cl2 | brak wpływu | 
Poniższy algorytm może być użyty do wyliczenia zależności temperaturowej elementu pomiarowego.
 K = [A+a(T - 25)] Σ bi . T i
 K= wartość skorygowana
 A= sygnał wyjściowy (0...100%)
 T= temperatura(ºC)
 bi= 6*10-4
 a= 0,04 (dla T≥25ºC)
 a= 0 ( dla T<25ºC)
 b0= 0,98125
 b2= 6*10-6




¿Estás interesado en este producto? ¿Necesita información adicional o precios individuales?
Usted debe estar conectado
|  Technical Data: | |
| Relative Humidity | |
| - measuring range | 0 ... 100 % rh | 
| - resolution | 0,5 %rh | 
| - accuracy | ±2 %rh 05„.35°C 25...85 %rh) | 
| ±3 %rh (15...35°C, 5...25 %rh and 85...95 %rh) | |
| - Additional error | ±0,1 % rh/K @t35°C | 
| Temperature | |
| - measuring ranges | see table | 
| - resolution | 0.1°C | 
| -accuracy | 0.5 K @ 25°C ±1 digit | 
| digit(see diagram) | |
| General: | |
| - operating voltage | 5...30 V DC | 
| - output signal: | |
| calHT01/1, calHT01/2 | 2 x 0...1 V (przy RL 3 1 MΩ) | 
| calHT 06/1, calHT 06/2 | digit (compatible with I2C) | 
| - storage temperature | - 65 ... 150°C | 
| - working temperature | - 40 ... 85°C | 
| - power consumption | < 1 mA | 
| Type | Measuring range [°C] | Labelling | 
| calHT01-0100(.R) | 0...100 | Green | 
| calHT01-3070(.R) | -30...70 | Black | 
| calHT01-2080(.R) | -20..80 | Red | 
| calHT01-4060(.R) | -40...60 | Yellow | 
| calHT06-4085(.R) | -40...85 | White | 


|  |  | 
| Temperature accuracy of calHT 01 and calHT 06 series | Temperature accuracy of calHT 01 and calHT 06 series | 

| S | 1 | 0 | 0 | 1 | 0 | 0 | 0 | 0 | A | 0 | 0 | 0 | 0 | C 4 | C 3 | C 2 | C 1 | A | P | 
| R/W | |||||||||||||||||||
 Microcontroller requests data byte according control command and reads out
| S | 1 | 0 | 0 | 1 | 0 | 0 | 0 | 1 | A | D 7 | D 6 | D 5 | D 4 | D 3 | D 2 | D 1 | D 0 | N A | P | 
| S Start | R/W | ||||||||
| P Stop | |||||||||
| A Acknowledge | from sensor | ||||||||
| N/A No Acknowledge | |||||||||
| C 4 | C 3 | C 2 | C 1 | |
|---|---|---|---|---|
| 0 | 0 | 0 | 0 | Read-Out Humidity | 
| 0 | 0 | 1 | 0 | Read-Out Temperature Byte 1 | 
| 0 | 0 | 1 | 1 | Read-Out Temperature Byte 2 | 
| D 7 | D 6 | D | D 4 | D 3 | D 2 | D 1 | D 0 | |
|---|---|---|---|---|---|---|---|---|
| 0/1 | 0/1 | 0/1 | 0/1 | 0/1 | 0/1 | 0/1 | 0/1 | Humidity | 
| 0/1 | 0/1 | 0/1 | 0/1 | 0/1 | 0/1 | 0/1 | 0/1 | Temperature Byte 1 | 
| X | X | X | X | 0/1 | 0/1 | 0/1 | 0/1 | Temperature Byte 2 | 
| Symbol | Parameter | min | max | jednostka | 
|---|---|---|---|---|
| tBUF | idle period between BUS actions | 4,7 | μs | |
| tD:SU | data set-up time | 250 | ns | |
| tD:HD | data hold time | 50 | ns | |
| tST:HD | start hold time | 4 | μs | |
| tST:SU | start set-up time | 4,7 | μs | |
| tSCL:L | SCL„low" time | 4,7 | μs | |
| tSCL:H | SCL „high" time | 4 | μs | |
| fSCL | SCL frequency | 100 | kHz | |
| t r | SDA, SCL LOW/HIGH time | 1 | μs | |
| tf | SDA, SCL HIGH/LOW time | 0,3 | μs | |
| tSTO:SU | stop set-up time | 4 | μs | |
| tSP | pulse width of spikes to be suppressed by input filter | 100 | ns | |
| CL | capacitive load SDA, SCL BUS (internal pull-up 120 kW) | 10 | pF | |
| tMUPD | update for measurement values | 40 | ms | |
| tSMPL | sampling period DA converter | 2 | 6 | ms | 

Instrukcja użytkowania
 Nigdy nie wolno dotykać aktywnej części bardzo czułego czujnika. Należy używać bardzo niskiego napięcia podczas lutowania elementu (temperatura lutowania 240°C, czas lutowania 2 sekundy). Należy oczyścić element pomiarowy delikatnie zdmuchując wszelkie pozostałości z montażu, można również przepłukać element w wodzie destylowanej. Nie wpływa to na charakterystykę czujnika, jednak dokładne pomiary mogą być dokonywane dopiero po całkowitym wysuszeniu elementu. Dotyczy to również sytuacji w której powstanie na elemencie rosa.
| Typ | FE 09/1, FE 09.R/1 FE 09/2, FE09.R/2 | FE 09/4 FE 09.R/4 | 
|---|---|---|
| Zakres działania | 0...100%rh | 0...100%rh | 
| Temperatura działania FE09/X | -60...200°C | -60...200°C | 
| Temperatura działania FE09.R/X | -40...110°C | -40...110°C | 
| Pojemność podstawowa | 135±10pF | 115±15pF | 
| Zmiana pojemności | 0,3±0,05pF/%rh | 0,27±0,08pF/%rh | 
| Dopuszczalne napięcie | maks. 3V bez DC | maks. 3V bez DC | 
| Częstotliwość pomiarowa | 5...200kHz | 5...200kHz | 
| Współczynnik strat | <0,03 (przy 10kHz) | <0,03 (przy 50kHz) | 
| Czas reakcji (T90) | 10s | 10s | 
| Histereza (MR 5...95%rh) | <1,5%rh | <1,5%rh | 
| Liniowość (MR 5...95%rh) | <1,5%rh | <1,5%rh | 
| Stabilność długoterminowa | <1%rh/a | <1%rh/a | 



| Substancja robocza | stężenie | 
|---|---|
| NH3 (tylko na krótki czas) | 100 ppm (2 x MAK) | 
| N2S | 20 ppm (2 x MAK) | 
| SO2 | 3,7 ppm (10 x MAK) | 
| No2 | 1,0 ppm (10 x MAK) | 
| O2 | 0,6 ppm (10 x MAK) | 
| Cl2 | brak wpływu | 
Poniższy algorytm może być użyty do wyliczenia zależności temperaturowej elementu pomiarowego.
 K = [A+a(T - 25)] Σ bi . T i
 K= wartość skorygowana
 A= sygnał wyjściowy (0...100%)
 T= temperatura(ºC)
 bi= 6*10-4
 a= 0,04 (dla T≥25ºC)
 a= 0 ( dla T<25ºC)
 b0= 0,98125
 b2= 6*10-6