PSUG-20T1 is an advanced surge simulator designed for rigorous electromagnetic (EMC) surge (SUG) immunity testing. This device is used to evaluate the performance of power cables and internal connections in electrical and electronic devices that may be exposed to high-energy transient interference from circuit switches or natural lightning strikes.
Parameter | Value |
---|---|
Model | PSUG-20T1 |
Open circuit voltage | |
Voltage Resolution | 100 V |
Rise Time | 1.2 µs 30% |
Duration | 50 µs 20% |
Short Circuit Current | 1000 A ~ 10000 A ±10% |
Rise Time | 8 µs ±20% |
Duration | 20 µs ±20% |
Output Polarity | +, -, +/-, -/+ |
Output Impedance | 2 Ω |
Step Voltage Range | 0 ~ 20000V |
Time Interval Between Pulses | 5 s ~ 9999 s |
Number of Pulses Test | 1c ~ 60000c |
Trigger mode | Manual, external, interval |
Phase synchronization | 0° ~ 359° (50 Hz or 60 Hz) (Synchronous/Asynchronous) |
Parameter | Value |
---|---|
CDN type | Built-in CDN |
Coupling | Single-phase, three-wire coupling |
Decoupling load | UAC MAX 240V/16A, UDC MAX 220V/16A |
Coupling capacitor | 9 μF (L-PE), 18 μF (L-L) |
Resistor Coupling | 12 Ω, 2 Ω |
Residual voltage at the input of the device under test | Not more than 15% of the applied surge voltage or twice the peak value of the rated coupling/decoupling network |
Érdekel ez a termék? További információra vagy egyedi árajánlatra van szüksége?
Musisz być zalogowany/a
PSUG-20T1 is an advanced surge simulator designed for rigorous electromagnetic (EMC) surge (SUG) immunity testing. This device is used to evaluate the performance of power cables and internal connections in electrical and electronic devices that may be exposed to high-energy transient interference from circuit switches or natural lightning strikes.
Parameter | Value |
---|---|
Model | PSUG-20T1 |
Open circuit voltage | |
Voltage Resolution | 100 V |
Rise Time | 1.2 µs 30% |
Duration | 50 µs 20% |
Short Circuit Current | 1000 A ~ 10000 A ±10% |
Rise Time | 8 µs ±20% |
Duration | 20 µs ±20% |
Output Polarity | +, -, +/-, -/+ |
Output Impedance | 2 Ω |
Step Voltage Range | 0 ~ 20000V |
Time Interval Between Pulses | 5 s ~ 9999 s |
Number of Pulses Test | 1c ~ 60000c |
Trigger mode | Manual, external, interval |
Phase synchronization | 0° ~ 359° (50 Hz or 60 Hz) (Synchronous/Asynchronous) |
Parameter | Value |
---|---|
CDN type | Built-in CDN |
Coupling | Single-phase, three-wire coupling |
Decoupling load | UAC MAX 240V/16A, UDC MAX 220V/16A |
Coupling capacitor | 9 μF (L-PE), 18 μF (L-L) |
Resistor Coupling | 12 Ω, 2 Ω |
Residual voltage at the input of the device under test | Not more than 15% of the applied surge voltage or twice the peak value of the rated coupling/decoupling network |